Jump to page №

Conductivity and Turbidity

TRUBOMAT GAB In-line turbidity system

TRUBOMAT GABIn-line turbidity system

TRUBOMAT GAB
In-line turbidity system

TRUBOMAT GAB measures the turbidity of liquid applying nephelometric method. For this purpose, a combined transmitted and scattered beams measurements are used, in which a transmitter and a receiver face each other and another transmitter is arranged orthogonally (at an angle of 90°). The atenuated and scattered lights are recorded by the electronic module (in the connection head) and process the results to transmit a linear output signal. TRUBOMAT GAB delivers a 4-20 mA signal, according to the set measuring range.

  • Continuous turbidity measurement according to ISO 7027
  • 90° scattered light measurement
  • 180° transmitted light measurement
  • Measuring cell: AISI 316 L
  • Measuring ranges:
  • 0.01 ... 20 FNU
  • 0.1 ... 500 FNU & 100 ... 1000 FAU
  • Output signal: 4-20 mA (4-wire)
  • Process connections: DIN flanges or Triclamp